Operator’s Manual
Charged Plate Analyzer
Model 268A-1
100 Housel Ave
PO Box 535 Lyndonville NY 14098 1-800-821-6001
585-765-2254
fax: 585-765-9330
monroe-electronics.com
Vue de la page 0
1 2 3 4 5 6 ... 11 12

Résumé du contenu

Page 1

Operator’s Manual Charged Plate Analyzer Model 268A-1 100 Housel Ave ⏐ PO Box 535 Lyndonville NY 14098

Page 2 - WARRANTY

APPENDIX I MODEL 268A-1T CHARGED PLATE ANALYZER WITH VARIABLE TIMER OPTION VARIABLE TIMER OPTION The Model 268A-1T Charged Plate Monitor with varia

Page 3 - TABLE OF CONTENTS

APPENDIX II Application Note APNE-0017 Plate Self Decay of Charged-Plate Monitors Introduction Charged-Plate Monitors (CPM’s) evaluate the ability

Page 4 - SPECIFICATIONS

The ESD Association Standard Test Method STM3.1 recognizes that a small amount of self decay may be unavoidable. This test method allows a maximum s

Page 5 - GENERAL INFORMATION

WARRANTY Monroe Electronics, Inc., warrants to the Owners, this instrument to be free from defects in material and workmanship for a period of two y

Page 6 - IMPORTANT NOTE

TABLE OF CONTENTS Warranty Section 1 ... Specifications Section 2 ...General Information Section 3 ...

Page 7 - OPERATION

Section 1 SPECIFICATIONS CHARGED PLATE/PROBE ASSEMBLY:  Designed to EOS/ESD Association standard  Standard tripod mount  Dimensions: 6” X 6”  C

Page 8

Section 2 GENERAL INFORMATION The Charged Plate Analyzer Model 268A separates into two components connected by a ten-foot cable. The top part wit

Page 9 - MAINTENANCE

IMPORTANT NOTE In order to maintain proper operation of the floating plate, it is essential that the support insulators be kept scrupulously clean.

Page 10 - APPENDIX I

Section 3 OPERATION BASIC OPERATION - DISCHARGE MODE: 1. Separate sections and position as necessary. Connect to the proper power source and pr

Page 11 - APPENDIX II

BASIC OPERATION - PEAK MODE: Peak mode is useful for checking balance in pulse ionizing systems. The timer is not associated with these measuremen

Page 12

Section 4 MAINTENANCE GENERAL: Field maintenance of Model 268A should be limited to cleaning of the insulating standoffs, cleaning of the case and