Literature References
Monroe Electronics recommends that anyone using Model 292 read the following standards from the
ESD Association:
EOS/ESD-ADV1.0 – Glossary of Terms
EOS/ESD-S4.1 – Work surfaces
EOS/ESD-S6.1 – Grounding
ESD-S7.1 – Floor Materials
EOS/ESD-S11.11 – Surface Resistivity
These documents are available directly from the ESD Association
7900 Turin Road
Suite 4
Rome NY 13440-2069
(315) 339-6937
General Guidelines for Testing
 Use both 5-pound electrodes for RTT (Resistance between two points) testing.
 Use one 5-pound electrode and one lead to a groundable point for RTG (Resistance to
Ground) testing.
 Use the optional resistivity adapter sled (removing leads and electrodes) for Resistivity
measurements.
 Make sure that the item being measured is electrically isolated (i.e. placed on an insulative
surface) as the meter will measure the lower resistance path.
 Make sure that the test leads are separated or the meter may measure the lower resistance
path
 When using the 5-pound electrodes:
o Place the electrodes no closer than 2 inches from edge of surface being measured.
o Place the electrodes no closer than 3 inches to any groundable point
o Preferred placements include: most commonly used surface portion, most worn,
center, and furthest from a groundable point.
o For RTG, connect the sensing lead with shielded plug to the groundable point.
 If the surface has sections (like tile flooring or garment panels), for RTT place the electrodes
on different sections.