Monroe-electronics Charge Plate - Monitor - model 288B Manuel d'utilisateur Page 4

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 22
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 3
4
Section 1
Specifications
All specifications are referred to plate voltage unless otherwise specified.
Display 240 x 64 character/graphic
Voltage - 3½ digit display (Decay and Peak reading)
Accuracy ±0.1% of reading ± 3 lsd
Resolution 1 volt for readings > 99 volts
0.1 volt for reading < 100 volts
Time - 4 digit display
Accuracy 0.1% of reading ± 1 lsd
Resolution 0.1 second for readings < 1000 seconds
1 second for reading > 999 seconds
Electrometer
Dynamic range ±1200 volts
Follower error < 10 mV
Speed of
Response < 10 msec for 1 kV to 0 volts (90% - 10%)
Bandwidth -3db @ 1Khz 20V
p-p
-3db @ 10Hz 2000V
p-p
Noise <12 mV
rms
Monitor output Divide by 200
Accuracy 0.1% of reading ± 12 mV
Output
Impedance 1 kΩ
Start Voltages 1000 volts ±0.3% Standard
Range ±10 to ±1000 volts.
Resolution settable to 1 volt
Accuracy 0.3% of setting ± 2.5 volts
Stop Voltages 100 volts ±3% Standard
Range 0 - ±995 volts
Resolution settable to 1 volt
Accuracy 0.3% of setting ± 2.5 volts
Charge voltage
Range 10 to 100 volts above the start voltage
Resolution settable to 1 volt increments
Accuracy 0.3% of setting ± 2.5 volts
Vue de la page 3