Monroe-electronics NanoCoulomb Meter - model 284 Manuel d'utilisateur Page 6

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 7
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 5
Page 4 284.CBSM-12/98-RCA
Section 4
CALIBRATION CHECK
The purpose of this section is to provide the operator of the Monroe Electronics
Model 284 Nanocoulomb Meter with a means of verifying that overall calibration
of the instrument is within the specified limits (±2%).
Method 1:
Procure a calibration capacitor. It should be a polystyrene or polypropylene
capacitor with a precisely known value of approximately 10nF (0.01µF) (Mallory
SX-110 or equivalent).
With the capacitor charged to exactly 1.00 volt and then discharged directly into
the input terminal (BNC connector) or between the inner and outer parts of the
Faraday cup, a meter reading equal to the capacitor value (±2%) should appear
on the Digital Panel Meter (DPM) set to the 20-nCoulomb range.
Charging the calibration capacitor to 10.00 volts and then discharging it into the
input should produce a DPM reading of ten times the capacitor value (±2%) on
the 200-nCoulomb range.
It is important to handle the calibration capacitor very carefully. Do not
hold the
capacitor by its body or discharge it by touching both leads with the fingers. Hold
the capacitor by one lead only
. Use a clip-lead connected between the GND jack
and this lead of the capacitor to maneuver the other lead of the capacitor
between the “hot” side of the charging source and the input terminal.
Method 2:
Use a secondary standard charge source such as a Keithley Model 263
Calibrator/Source, following manufacturer’s instructions.
Incorrect DPM readings indicate a need for calibration.
Vue de la page 5
1 2 3 4 5 6 7

Commentaires sur ces manuels

Pas de commentaire