Monroe-electronics Electrostatic Fieldmeter - Static Monitor - model Manuel d'utilisateur Page 52

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 96
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 51
51
Convergent field lines such, as these tend to “enhance” the measurement and cause the apparent field strength
to increase, thus producing a high reading. Although this may be relatively inconsequential where a direct
measurement of surface voltage is desired — for example, where a probe whose specified sensitivity is 10kV/cm
is being used at 1cm to produce an indicated full scale value of up to 10,000 volts — given a large enough
surface and a properly sized shroud, that same probe could be used to produce a measurement of surface
voltage up to 100,000 volts at 10cm with fairly good accuracy.
Tests performed in our laboratory produced the following:
Some conclusions may be drawn from the data shown in Figure 17 :
1. An "F” probe produces a reading about 12% high with no shroud at 1" (25mm) and about double at 4"
(100mm).
2. When provided with a 6" diameter shroud, the error drops to fewer than 10% at spacings of up to
almost 4".
3. A shroud is unnecessary on an "E" probe at 1" or less and would produce the same improvement as
with the "F" probe at greater spacing.
4. As a general rule of thumb, a shroud (or the size of the face of the probe) should be about times
the spacing to keep errors to about 10% or less.
0.8
1
1.2
1.4
1.6
1.8
2
0 2 4 6
Inches from Target
Enhancement Factor
"F" probe without shroud
"E" probe
without shroud
"F" probe with
6" dia. shroud
1036 DATA
Figure 17
Vue de la page 51
1 2 ... 47 48 49 50 51 52 53 54 55 56 57 ... 95 96

Commentaires sur ces manuels

Pas de commentaire