APNE-0016
84
Probe Model Sensitivity V
HV
V
IND
1036_-2 ±100 V/cm 254 V 100.0 V
1036_-3 ±1 kV/cm 2540 V 1.000 kV
1036_-4 ±10 kV/cm 2540 V 1.00 kV
1036_-5 ±20 kV/cm 2540 V 0.50 kV
1036_-6 ±10 kV/in 1000 V 1.000 kV
1036_-7 ±1 kV/in 1000 V 1.00 kV
Table II-2
HV Source and Meter Reading for Probe Standardization
Intrinsic Safety (IS) Barriers
Model 1036E and 1036F Electrostatic Fieldmeter Probes meet Factory Mutual Research
Corporation requirements for Class I, Division 1, Groups C and D hazardous locations when
installed in accordance with the appropriate Monroe Electronics, Inc. control drawings. To
comply, approved safety barriers must be used as shown in these control drawings. These
drawings are included in Appendix B – Intrinsic Safety Barriers, in the Model 177A
Instruction Manual that is shipped with the instrument, or are available from Monroe
Electronics.
Copies of the Factory Mutual Research Corporation report 1Q3A9.AX, specific to these
probes, are available on request from:
Monroe Electronics, Inc.
100 Housel Avenue
Lyndonville, NY 14098
Tel. (585) 765-2254,
Fax. (585) 765-9330